Skip to main content

The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is not authorized for use with thoracoabdominal and pararenal branched devices indicated with the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis as a branch component in Canada.

Holden A, Merrilees S, Buckley B, Connor B, Colgan F, Hill A. First-in-human experience with the Gore Balloon-Expandable Covered Endoprosthesis in iliac artery occlusive disease. Journal of Endovascular Therapy 2017;24(1):11-18. 

Abstract 

Purpose: 

To report the first-in-human iliac artery experience of a new balloon-expandable covered endoprosthesis.  

Methods:

A prospective, single-center pilot study recruited 30 symptomatic patients (mean age 64 years; 18 men) to evaluate the safety and early efficacy of the new Gore balloon-expandable covered endoprosthesis for the treatment of de novo or restenotic common and/or external iliac artery lesions. According to protocol, up to 2 discrete lesions could be treated with a maximum total treated length ≤110 mm. Follow-up included clinical evaluation with duplex ultrasound at 1, 6, and 12 months. Data are presented through 12-month follow-up. The primary safety endpoint was a composite of device- or procedure-related death, myocardial infarction, or amputation in the treated leg within 30 days of the index procedure. Multiple performance outcomes were also evaluated.  

Results:

The primary 30-day safety endpoint was 0%. Per-subject estimates of primary patency, freedom from target lesion revascularization, and freedom from target vessel revascularization were 100% at 1 and 6 months and 96.6% at 12 months. Estimates of assisted primary and secondary patency were both 100% at 12 months. Freedom from major adverse events at 12 months was 100%. Most patients experienced improvements in Rutherford category, ankle-brachial index, and functional status that were sustained to 12 months. 

Conclusion: 

This positive first-in-human experience with the Gore balloon-expandable covered endoprosthesis suggests this device will have an important role in the management of aortoiliac occlusive disease.


IFU Consult instructions

Refer to Instructions for Use at eifu.goremedical.com for a complete description of all applicable indications, warnings, precautions and contraindications for the markets where this product is available. RXOnly

INDICATIONS FOR USE IN THE U.S.: The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is indicated for the treatment of de novo or restenotic lesions found in iliac arteries with reference vessel diameters ranging from 5 mm–13 mm and lesion lengths up to 110 mm, including lesions at the aortic bifurcation. The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is also indicated for use with thoracoabdominal and pararenal branched devices indicated with the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis as a branch component.*

CONTRAINDICATIONS: Do not use the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis in patients with known hypersensitivity to heparin, including those patients who have had a previous incident of Heparin-Induced Thrombocytopenia (HIT) type II.  

* Not applicable to Reduced Profile GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis. (BXB catalogue numbers).