Pivotal Study of a Next-Generation Balloon-Expandable Stent-Graft for Treatment of Iliac Occlusive Disease
The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is not authorized for use with thoracoabdominal and pararenal branched devices indicated with the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis as a branch component in Canada.
Bismuth J, Gray BH, Holden A, Metzger C, Panneton J; VBX FLEX Study Investigators. Pivotal study of a next-generation balloon-expandable stent-graft for treatment of iliac occlusive disease. Journal of Endovascular Therapy 2017;24(5):629-637.
Abstract
Purpose:
To evaluate the safety and efficacy of a next-generation balloon-expandable stent-graft for the treatment of arterial occlusive disease in patients with de novo or restenotic lesions in the common and/or external iliac arteries.
Methods:
A prospective, multicenter, single-arm safety and efficacy study (ClinicalTrials.gov identifier: NCT02080871) of the VBX Stent Graft for treatment of de novo or restenotic lesions in the iliac arteries was conducted under an Investigational Device Exemption at 26 US sites and 1 New Zealand center. The study eligibility criteria were established with the aim of enrolling more “real-world” patients compared with previous iliac stent studies supporting Food and Drug Administration approval. The study enrolled 134 patients (mean age 66±9.5 years; 79 men) with 213 iliac lesions. Three-quarters of the population (101, 75.4%) had Rutherford category 3 ischemia, and 43 (32.1%) patients presented with TASC II C/D lesions. The primary endpoint was a composite of device- or procedure-related death within 30 days, myocardial infarction within 30 days, target lesion revascularization (TLR) within 9 months, and amputation above the metatarsals in the treated leg within 9 months.
Results:
In all, 234 devices were implanted in 213 lesions with 100% technical success; 57 (42.5%) patients received kissing stents at the aortic bifurcation. No devices exhibited a discernable change in length after final deployment as determined by independent core laboratory quantitative angiographic analysis. At 9 months, 3 (2.3%) of 132 patients (1 lost to follow-up, 1 unrelated death) experienced a major adverse event (3 TLRs) related to the primary endpoint. At 9 months, there were no device-related serious adverse events or unanticipated adverse device effects.
Conclusions
The next-generation balloon-expandable stent-graft demonstrated notable 9-month safety and efficacy in treating iliac occlusive disease in patients reflecting common clinical practice.
Refer to Instructions for Use at eifu.goremedical.com for a complete description of all applicable indications, warnings, precautions and contraindications for the markets where this product is available. RXOnly
INDICATIONS FOR USE IN THE U.S.: The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is indicated for the treatment of de novo or restenotic lesions found in iliac arteries with reference vessel diameters ranging from 5 mm–13 mm and lesion lengths up to 110 mm, including lesions at the aortic bifurcation. The GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis is also indicated for use with thoracoabdominal and pararenal branched devices indicated with the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis as a branch component.*
CONTRAINDICATIONS: Do not use the GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis in patients with known hypersensitivity to heparin, including those patients who have had a previous incident of Heparin-Induced Thrombocytopenia (HIT) type II.
* Not applicable to Reduced Profile GORE® VIABAHN® VBX Balloon Expandable Endoprosthesis. (BXB catalogue numbers).